TECHNICAL SPECIFICATION |
|
Infusion mode: volumetric |
|
Pumping mechanism: peristaltic finger-shape pump, detachable mechanism for cleaning |
|
Infuse rate unit: d/min, mL/h and volume/time |
|
IV set: standard IV sets with diameter 3.4 – 4.5 mm |
|
Flow rate range: 0.1 – 1200 mL/h |
|
Increment: 0.1 mL/h (between 0.1 mL/h and 99.9 mL/h); 1mL/h (between 100 mL/h and 1200 mL/h) |
|
One point calibration to keep the accuracy at any speed |
|
Flow rate accuracy: ≤ ±5% (after calibration) |
|
Volume limit: 1 – 9999 mL |
|
Volume infused display: 0 – 9999 mL |
|
Purge rate: 600 mL/h |
|
KVO rate: 1 mL/h |
|
Bolus mode: available for bolus mode, flow rate adjustable |
|
Occlusion pressure: pressure detected and the device with 8 sensitivities setting, 40-140 kPa |
|
Air-in-line detection: air bubble detected by ultrasonic and the device with 8 sensitivities setting, 50 – 1000 μL |
|
Alarm volume: 8 levels setting |
|
Anti reverse function: anti-reverse function on motor to prevent upstream |
|
IV set management and flow rate calibration: IV sets can be added, stored and deleted in memory and easily used after calibration, totally 20 brands storage |
|
Self-check: self check when start up and during infusion |
|
Audio and visual alarm: air-in-line, door open, occlusion, infusion finished, pause overtime, low battery, battery damaged, abnormity of motor, sound, pressure, sensor, ultrasonic sensor and AC power supply. |
|
Display: blue LCD with high brightness, dynamical pressure display |