 TECHNICAL SPECIFICATION  | 
 
  | 
| 
 Infusion mode: volumetric  | 
 
  | 
| 
 Pumping mechanism: peristaltic finger-shape pump, detachable mechanism for cleaning  | 
 
  | 
| 
 Infuse rate unit: d/min, mL/h and volume/time  | 
 
  | 
| 
 IV set: standard IV sets with diameter 3.4 – 4.5 mm  | 
 
  | 
| 
 Flow rate range: 0.1 – 1200 mL/h  | 
 
  | 
| 
 Increment: 0.1 mL/h (between 0.1 mL/h and 99.9 mL/h); 1mL/h (between 100 mL/h and 1200 mL/h)  | 
 
  | 
| 
 One point calibration to keep the accuracy at any speed  | 
 
  | 
| 
 Flow rate accuracy: ≤ ±5% (after calibration)  | 
 
  | 
| 
 Volume limit: 1 – 9999 mL  | 
 
  | 
| 
 Volume infused display: 0 – 9999 mL  | 
 
  | 
| 
 Purge rate: 600 mL/h  | 
 
  | 
| 
 KVO rate: 1 mL/h  | 
 
  | 
| 
 Bolus mode: available for bolus mode, flow rate adjustable  | 
 
  | 
| 
 
 Occlusion pressure: pressure detected and the device with 8 sensitivities setting, 40-140 kPa  | 
 
  | 
| 
 Air-in-line detection: air bubble detected by ultrasonic and the device with 8 sensitivities setting, 50 – 1000 μL  | 
 
  | 
| 
 
 
 
 
 
 
 
 
 
 
 
 Alarm volume: 8 levels setting  | 
 
  | 
| 
 Anti reverse function: anti-reverse function on motor to prevent upstream  | 
 
  | 
| 
 IV set management and flow rate calibration: IV sets can be added, stored and deleted in memory and easily used after calibration, totally 20 brands storage  | 
 
  | 
| 
 Self-check: self check when start up and during infusion  | 
 
  | 
| 
 Audio and visual alarm: air-in-line, door open, occlusion, infusion finished, pause overtime, low battery, battery damaged, abnormity of motor, sound, pressure, sensor, ultrasonic sensor and AC power supply.  | 
 
  | 
| 
 Display: blue LCD with high brightness, dynamical pressure display  | 
|